Checking a thermometer sounds easy until you actually have to do it. An ice bath can be set up, but only for a single point and only if you do it properly. Boiling water provides another point, though it varies with atmospheric pressure. If you want to check −20 °C or +220 °C, you suddenly need a calibration bath or a dry-block calibrator—meaning equipment that costs more than the thermometers being tested. As a result, checks are often postponed—and an unchecked thermometer is, in practice, a thermometer left to wishful thinking.
What It Is
The ETI MicroCheck is a calibrated temperature simulator for Type K thermocouple thermometers. This model simulates three fixed points: −20 °C, 0 °C, and +220 °C. The instrument features a 10 mm LCD with an alphanumeric line, as well as selectable parameters for °C/°F, auto-power off, and internal or external CJC. It is supplied with a one-metre Type K lead with miniature connectors, a protective boot, and a 3-point UKAS Calibration Certificate indicating the deviation at each point. The range is available in four variants with different points: MicroCheck 1 (0/100/500 °C), 2 (−20/20/200 °C), 3 (−20/0/220 °C), and 4 (−20/0/100 °C).
The Problem It Solves
A thermocouple works because two dissimilar metals generate a small voltage that depends on the temperature difference between the measurement tip and the junction. The thermometer reads this voltage—a few tens of microvolts per degree—and converts it into a temperature reading.
The simulator leverages precisely this: instead of generating a temperature, it generates the exact voltage that a Type K thermocouple would produce at a given temperature. Plug it into the thermometer’s input, select the test point, and read what the instrument displays. No heating up, no waiting for stabilization, and no risk of non-homogenous test media. Three known points can be verified in minutes, without any specialized equipment.
The selectable CJC (cold junction compensation) setting ties directly into this. The thermometer must compensate for the temperature at its own input terminals to calculate the voltage correctly, and this is one of the things that can actually fail within an instrument. With the simulator, you can verify that this compensation is functioning as intended.
And here is the key limitation, which is important to clarify: a simulator tests the thermometer—the electronics, the calculation, the display—but not the sensor. The probe is unplugged during the test. This means that a passed simulator test tells you nothing about a probe that is bent, has a damaged tip, or has suffered from drift due to aging. In practice, the probe is also the component exposed to the most wear and tear.
The takeaway is not that this method is inferior, but that it addresses a specific question: is the instrument calculating correctly? This question can be asked frequently, cost-effectively, and without taking equipment out of service. Probe verification can then be performed separately and less frequently—using an ice bath, reference thermometer, or external calibration—and together, both methods cover the entire measurement chain.
Three Typical Use Cases
- Routine in-house checks: monthly or quarterly verification of your facility’s Type K thermometers, documented against a traceable certificate.
- Troubleshooting deviations: when a reading seems off, plug in the simulator and determine within a minute whether the issue lies in the instrument or elsewhere.
- Incoming inspection: fast functional checks of newly purchased or newly repaired instruments before placing them into service.
Why It Pays Off
Compare this to external calibration for every single instrument. Shipping costs, service fees, and lead times per thermometer mean that, in practice, calibration intervals stretch longer than they should, leaving you without your instrument in the meantime. Having an in-house simulator turns verification into a two-minute task that can be carried out at any time on all instruments—meaning it actually gets done.
The real value lies in limiting retrospective risk. If a thermometer is found to be reading incorrectly, the immediate question is always: for how long? With documented in-house checks, you know the fault occurred since the last check—days or weeks ago—rather than having the uncertainty trace all the way back to the last external calibration. That is the difference between quarantining a couple of batches and having to write off an entire quarter.